Abstract

A novel high-performance liquid chromatographic (HPLC) method for the quantification of diclofenac in human plasma was set up. Samples, added with ibuprofen (used as internal standard) were purified by solid-phase extraction using Abselut Nexus cartridges (Varian) not requiring pre-conditioning. Drugs of interest were eluted directly into the autosampler vials and injected. The recovery of diclofenac was 92%, the analysis lasted 7 min with a sensitivity of 5 ng/ml and intra- and inter-day RSDs of 3 and 8%, respectively. The pharmacokinetics of diclofenac after oral and rectal administration in 10 healthy volunteers are reported.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.