Abstract

A sensitive and specific high-performance liquid chromatographic (HPLC) assay was developed for the quantitation of mitoxantrone in plasma using electrochemical detection. Bisantrene was chosen as the internal standard. A reversed-phase, 10-μm μBondapak C 18 analytical column (30 cm × 3.9 mm) with an isocratic mobile phase of 28% acetonitrile in 80 m M sodium formate buffer (pH 3.0) was used. The eluent was monitored by both electrochemical detection at an applied potential of +0.75 V vs. Ag/AgCl and visible absorbance at 660 nm. Only electrochemical detection was able to quantitate the internal standard and provided ten times higher sensitivity than visible absorbance for mitoxantrone with a detection limit as low as 0.1 ng/ml. Calibration curves in the range 0.1–1000 ng/ml showed good linearity ( r = 0.998) and precision (coefficient of variation < 10%). This HPLC method utilized a reproducible and inexpensive liquid-liquid extraction procedure. Using methylene chloride, the extraction efficacy of mitoxantrone from plasma was 85.3% with a coefficient of variation less than 2.1%. This new assay was then applied to measure mitoxantrone concentrations in plasma obtained from two leukemic patients receiving 12 mg/m 2 mitoxantrone as a 1-h infusion.