Abstract

A simple, sensitive and accurate reverse phase high-performance liquid chromatographic (RP-HPLC) method with photo-diode array detector (PDA) was developed and validated for the determination of amphotericin B (AMB) in the rat plasma using a new internal standard (IS) α-naphthol. The plasma samples were subjected to protein precipitation with methanol prior to a HPLC analysis. Chromatographic separations were achieved on a Nucleosil ® 100-5C18 (150 mm × 4.6 mm) column. The mobile phase consisted of acetonitrile and sodium acetate buffer (pH 4; 10 mM) in a gradient mode. Detection was carried out at a wavelength of 407 and 294 nm for AMB and IS, respectively. The retention times of AMB and IS were about 6.8 and 7.8 min, respectively. The calibration curve was linear in the range of 10–2000 ng mL −1 for AMB ( r 2 > 0.998). No significant matrix effect was observed on quantification of AMB or IS. At three quality control concentrations of 20, 500, and 2000 ng mL −1, the intra-day and inter-day relative standard deviation ranged from 1.13% to 4.91%. The limit of detection (LOD) was 5 ng mL −1 and the limit of quantification (LOQ) was 10 ng mL −1 for AMB in rat plasma. This method is simple, sensitive, rapid and does not require any extensive sample purification before injecting into HPLC.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.