Abstract
The sensitivity and performance of an asymmetric Mach-Zehnder interferometer (aMZI) were compared to those of quartz crystal microbalance with dissipation (QCM-D). The binding of streptavidin to sensor chips coated with poly-l-lysine (PLL), modified with biotin and oligoethyleneglycol (OEG) (PLL-biotin), was used to compare the binding signals obtained from both technologies. PLL-biotin proved to be an efficient method to add bioreceptors to both the QCM-D and aMZI chips. The final, saturated value of streptavidin binding was compared with those from aMZI (253 ng cm-2) and QCM-D (460 ng cm-2). These values were then used to evaluate that 45% of the measured streptavidin mass in the QCM-D came from hydrodynamically coupled water. Importantly, the signal-to-noise ratio of the aMZI was found to be 200 times higher than that of the QCM-D. These results indicate the potential of the aMZI platform for highly sensitive and accurate biosensing applications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.