Abstract
A highly sensitive, rapid assay method has been developed and validated for the estimation of ropinirole (RPR) in human plasma with liquid chromatography coupled to tandem mass spectrometry with electrospray ionization in the positive-ion mode. A solid-phase process was used to extract RPR and citalopram (internal standard, IS) from human plasma. Chromatographic separation was operated with 0.2% ammonia solution:acetonitrile (20:80, v/v) at a flow rate of 0.50 mL/min on a Hypurity C(18) column with a total run time of 3.2 min. The MS/MS ion transitions monitored were 261.2 --> 114.2 for RPR and 325.1 --> 209.0 for IS. Method validation and clinical sample analysis were performed as per FDA guidelines and the results met the acceptance criteria. The lower limit of quantitation achieved was 3.45 pg/mL and the linearity was observed from 3.45 to 1200 pg/mL. The intra-day and inter-day precisions were in the range of 4.71-7.98 and 6.56-8.31%, respectively. This novel method has been applied to a pharmacokinetic study of RPR in humans.