Abstract

A highly sensitive method for the measurement of malondialdehyde as thiobarbituric acid—malondialdehyde complex by reversed-phase high-performance liquid chromatography with fluorescence detection in biological samples is described. As samples, 20 μl of rat plasma or 10% (w/v) liver homogenate mixed with 2.0% thiobarbituric acid in 2 M sodium acetate buffer containing 0.05% butyl hydroxytoluene (pH 3.5) were heated at 95°C for 45 min to give the complex. The complex, extracted with n-butanol, was chromatographed on a system equipped with a reversed-phase column, and the eluted peak was monitored with a fluorescence detector (excitation wavelength 515 nm, emission wavelength 553 nm). The mobile phase was a acetonitrile—water (2:8, v/v) under isocratic conditions at ambient temperature, and a single analysis was done in ca. 4 min. The minimum detection level for malondialdehyde was as low as 0.05 pmol. The n-butanol extract was stable at least for 3 days. The simple mobile phase, the extremely sensitive detection limit, and the stability of the complex make this system applicable to routine clinical analysis with a small amount of tissue or biopsy sample.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.