Abstract

Evidentiary traces may contain low quantities of DNA, and regularly incomplete short tandem repeat (STR) profiles are obtained. In this study, higher capillary electrophoresis injection settings were used to efficiently improve incomplete STR profiles generated from low-level DNA samples under standard polymerase chain reaction (PCR) conditions. The method involves capillary electrophoresis with higher injection voltage and extended injection time. STR peak heights increased six-fold. Inherent to the analysis of low-level DNA samples, we observed stochastic amplification artifacts, mainly in the form of allele dropout and heterozygous peak imbalance. Increased stutter ratios and allele drop-in were rarely seen. Upon STR typing of 10:1 admixed samples, the profile of the major component did not become overloaded when using higher injection settings as was observed upon elevated cycling. Thereby an improved profile of the minor component was obtained. For low-level DNA casework samples, we adhere to independent replication of the PCR amplification and boosted capillary electrophoresis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.