Abstract

A high-voltage probe measuring system with an overall rise time of less than 50 psec and a peak voltage capability of 20 kV has been developed. Probe perturbations at the 50 Ω impedance level are minimized by maintaining a moderately high input impedance, in conjunction with the absorption of internal reflection waves by means of a lossy distributed transmission line. System long-term stability and constructional reproducibility have been investigated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.