Abstract
We compared the impact of low and high-pressure balloon inflation on acute and late angiographic results of Multilink stent. Low-pressure balloon inflation (9.5 +/- 1.9 atm) was used in 43 stents and high pressure (17.1 +/- 1.5 atm) in 44. A larger immediate luminal gain was achieved in stents with high-pressure balloon inflation (1.80 +/- 0.26 vs. 1.47 +/- 0.62; P = 0.002), resulting in a larger mean diameter in this group (2.71 +/- 0.37 vs. 2.48 +/- 0.47; P = 0.017). At follow-up, a larger luminal diameter was achieved in the high pressure group (1.93 +/- 0.72 vs. 1.45 +/- 0.66; P = 0.002) and a trend to a lower rate of angiographic restenosis (15% vs. 38%, P = 0.08).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.