Abstract

Time-of-flight secondary ion mass spectrometry (ToF-SIMS) has become a promising analytical tool for molecular profiling in biological applications. However, its ultrahigh vacuum environment and matrix effects hamper the absolute quantitation of solution samples. Herein, we present a rapid high-throughput platform for quantitative ToF-SIMS analysis of amino acids in matrix deposits formed from freeze-dried solution drops through ice sublimation on a parylene film microarray substrate. Droplets of the amino acid solutions, which were mixed with stable isotope-labeled phenylalanine (F*) of high concentration (10 mM), were loaded on wells of the microarray, then frozen and evaporated slowly below the freezing point, forming continuous solid-phase F* matrix deposits. The amino acids (≤500 μM), adequately well dispersed throughout the F* matrix deposits on each well, were quantitatively analyzed by ToF-SIMS in a rapid and high-throughput fashion. The lower limit of quantitation reached below 10 μM.