Abstract

Surfactant degradation in biopharmaceuticals has recently gained significant attention in the pharmaceutical industry. Specifically, hydrolytic degradation of polysorbates, leading to the release of free fatty acids potentially forming visible particles, is a key theme in technical development. To address this emerging topic, we present the development of a fully automated liquid-chromatography single quad mass detector method for the quantification of free fatty acids in biopharmaceuticals. For the first time, we have quantified the longer chain fatty acid degradation products of polysorbate, palmitic and stearic acid, allowing reliable detection and early critical insights for process improvements. This high-throughput method was validated underlining its robust performance in an interlaboratory trial as well as high flexibility allowing different robotic platforms and preparation techniques. The combination of automated sample preparation, separation by liquid chromatography and single quad mass detection makes the validated fatty acid mass spectrometry assay ready for routine use in a regulated environment.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.