Abstract
The paper describes a high-temperature superconducting magnetic system (HTS SMS) to equip an experimental stand intended for neuron activity researches under constant and low-frequency magnetic fields up to 1 T. The design of the magnetic system together with its electromagnetic and cryogenic parameters is briefly discussed. The test results of the preliminary experiments conducted in liquid nitrogen at 77 K for two interchangeable magnets are given. The first magnet was manufactured in the form of a double pancake coil wound with 4 mm high HTS tape. The second magnet was made of pure copper wire with no frame and was impregnated with a thermally conducting epoxy resin. The advantages of the HTS pancake coil were demonstrated in comparison with the cryo-resistive solenoid. Low energy consumption of the HTS magnetic system will allow conducting continuous non-invasive monitoring of biological objects in a magnetic field. Keywords: superconductivity, high-temperature superconductor, REBCO tape, cryogen magnetic system, magnetic field, neuron activity.