Abstract
A pulsed electron spin resonance (ESR) microimaging system operating at the Q-band frequency range is presented. The system includes a pulsed ESR spectrometer, gradient drivers, and a unique high-sensitivity imaging probe. The pulsed gradient drivers are capable of producing peak currents ranging from ∼9 A for short 150 ns pulses up to more than 94 A for long 1400 ns gradient pulses. Under optimal conditions, the imaging probe provides spin sensitivity of ∼1.6 × 10(8) spins∕√Hz or ∼2.7 × 10(6) spins for 1 h of acquisition. This combination of high gradients and high spin sensitivity enables the acquisition of ESR images with a resolution down to ∼440 nm for a high spin concentration solid sample (∼10(8) spins∕μm(3)) and ∼6.7 μm for a low spin concentration liquid sample (∼6 × 10(5) spins/μm(3)). Potential applications of this system range from the imaging of point defects in crystals and semiconductors to measurements of oxygen concentration in biological samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.