Abstract

The performance and design of an imaging position sensitive, particle beam detector will be presented. The detector is minimally invasive, operates over a wide dynamic range (> 10 10), and exhibits high spatial resolution. The secondary electrons produced when a particle beam passes through a thin foil are imaged using stigmatic ion optics onto a two-dimensional imaging detector. Due to the low scattering cross section of the 6 nm carbon foil the detector is a minimal perturbation on the primary beam. A prototype detector with an image resolution of approximately 5 μm for a field of view of 1 mm has been reported [R.W. Odom, M.D. Strathman, S.E. Buttrill, Jr., and S.M. Bauman, Nucl. Instr. and Meth. B44 (1990) 465]. A higher resolution detector for imaging small beams (< 50 μm) with an image resolution of better than 0.5 μm has since been developed and its design is presented.