Abstract
We have studied the accuracy of blood volume measurements using carbon monoxide (CO)-labelled haemoglobin (COHb) injection and dilution (CO method) by comparing changes in red cell volume (RCV) measured using the CO method and 51Cr-labelled erythrocyte dilution (51Cr method) in a haemorrhage and infusion model in rabbits. RCV was measured repeatedly using the CO method at four different blood volume stages (stages I-IV). At stages I and IV, RCV was measured simultaneously using the 51Cr method. In comparing the sum of the circulating RCV and extracted RCV (SUM RCV) using the CO method, the values were almost equal and there were no significant differences between the values at the four stages. In comparing circulating RCV measured using the CO method and the 51Cr method, mean difference between the two methods was 0.80 (SD 0.76) ml kg-1 or 4.7 (4.6)%, and a positive correlation was observed (r = 0.91). We conclude that the CO method can be used to measure blood volume during perioperative periods in infants because it avoids use of a radioactive tracer, is simple and repeated measurements are possible.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.