Abstract

Abstract To solve the problem of calibrating the radius of a ball indenter in a hardness tester, a laser confocal radius measurement and calibration method for the ball indenter is proposed without separating the ball from the body of the indenter. The laser confocal radius measurement and calibration method uses the maximum of the confocal axial intensity curve to precisely identify the cat’s eye and confocal position of the test ball indenter. The distance between these two positions is then measured to achieve high-precision radius measurement. The theoretical analyses and experimental results indicate that the radius measurement uncertainty of the ball indenter with a diameter of 1.5875 mm is within 0.12 μm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.