Abstract

A rapid and selective reversed-phase high-performance liquid chromatographic assay with gradient elution and diode-array detection for diacetylmorphine, morphine, codeine, and their free and glucuronidated metabolites in plasma, was developed. After addition of ethylmorphine as internal standard the plasma samples were extracted using C 18 ODS-2 solid-phase columns with a recovery better than 80%. The limit of quantitation using an injection volume of 2 μl was 25 ng/ml for each compound. The intra- and inter-day precision was better than 5%. The described method cannot only be used for pharmacokinetic studies but also for intoxication cases to monitor a wide range of opiates.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.