Abstract
Methamphetamine in the human urine wasdirectly determined by high performance liquid chromatography (HPLC) by using a polyvinylalcohlic column and alkali mobile phase. The detailed conditions were as follows : column : Asahipak GS-320M 7.6 mm i.d.×100 mm, column temp. : 40°C, mobile phase : 80% {0.02 M NaHCO3+0.02 M Na2CO3 (8 : 2)} +20% CH3CN, flow rate : 1.0 ml/min, detector : UV photometer (207 nm). The urine was directly injected into HPLC. Methamphetamine was eluted by adsorption chromatography, and separated from endogenous urine constituents. Determination of methamphetamine was carried out successfully by using the calibration curve by peak high (1-10μg/ml). The recovery of methamphetamine added to the human urine (5μg/ml) was 92.0±1.0%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.