Abstract

This paper describes a sensitive HPLC-electrochemical detection analytical method for determining the concentration of the intravenous anesthetic, propofol, in human or rat plasma or serum and a variety of rat tissues. Internal standard and drug are extracted from serum or plasma and other tissues with pentane. 2,6- tert.-Butylmethylphenol is used as internal standard. It includes a novel steam distillation procedure for separating the highly lipophilic propofol from skin and fat. The plasma/serum assay has a precision of 1–4% (C.V.) in the range 10 ng/ml to 1 μg/ml and permits the assay of 5 ng/ml from 0.1 ml of plasma/serum. The tissue procedure allows the estimation of 50 ng/g in 0.1 g of tissue for most of the major organs with less than 2% (C.V.) precision. This assay was used to measure propofol concentrations in plasma/serum and tissue samples in support of a project to develop a physiological pharmacokinetic model for propofol in the rat.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.