Abstract

A sensitive and specific high-performance liquid chromatographic (HPLC) method with UV detection was developed for the determination of minocycline in human plasma and parotid saliva samples. Samples were extracted using an Oasis™ HLB cartridge and were injected into a C 8 Nucleosil column. The HPLC eluent contained acetonitrile–methanol–distilled water–0.1% trifluoroacetic acid (25:2:72.9:0.1, v/v). Demeclocycline was used as internal standard. The assay showed linearity in the tested range of 0.1–25 μg/ml. The limit of quantitation was 100 ng/ml. Recovery from plasma or parotid saliva averaged 95%. Precision expressed as %CV was in the range 0.2–17% (limit of quantitation). Accuracy ranged from 93 to 111%. In the two matrices studied at 20 and 4°C, rapid degradation of the drug occurred. Frozen at −30°C, this drug was stable for at least 2 months, the percent recovery averaged 90%. The method’s ability to quantify minocycline with precision, accuracy and sensitivity makes it useful in pharmacokinetic studies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.