Abstract

Eight ionic liquids (ILs) are subjected to both positive-ion and negative-ion direct analyses in real time (DART) Fourier transform ion cyclotron resonance (FT-ICR) mass spectrometry (MS). First, their ability to deliver evenly distributed cluster ion series covering a wide m/z range is explored. Then, one of the ILs exhibiting particularly useful cluster ion series in either ion polarity is applied for mass calibration. Using 1-butyl-3-methylimidazolium tricyanomethide delivers positive cluster ions suitable for mass calibration in the m/z 100-4,000 range and covers the m/z 100-2,000 range in negative-ion DART-MS. The corresponding mass reference lists are provided for either polarity. Furthermore, based on 1-butyl-3-methylimidazolium tricyanomethide, a high-mass record of m/z > 5,000 for positive-ion DART-MS is presented. The mass calibration procedure is finally validated by application to established standard compounds such as polydimethylsiloxanes, perfluorononanoic acid, and Ultramark 1621, a mixture of hexakis (fluoroalkoxy) phosphazenes. Further proof is presented by consistent exact mass differences between adjacent cluster ions.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.