Abstract

To solve the problem that the dew point water vapor content in a low-temperature wind tunnel spans six orders of magnitude and the dynamic range of dew point measurement is high, the dew point measurement technology of dualwavelength laser absorption spectrum is studied in this paper. Using the hitran2012 database, the dew point absorption spectrum is simulated. 1854nm laser is selected for high dew point measurement and 2626nm laser for low dew point measurement. Through the combination of two bands, the high dynamic dew point measurement in the range of - 100℃~30℃ is realized, which meets the requirements of the dew point measurement range of low-temperature wind tunnel, and the measurement error is less than 2℃.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.