The high-dose hook effect is a well-known phenomenon of two-site immunoassays including those for human chorionic gonadotrophin (hCG). We investigated the occurrence of a high-dose hook effect in six routinely available hCG assays using a sample with a total hCG concentration of approximately 3,600,000 IU/L. Dilutions of a sample with high hCG concentration were analysed using six common methods: Advia Centaur, Immulite 2000, Dimension RxL, Unicel DxI 800, Roche E170 and Abbott Architect. The measured concentrations and corresponding assay signals were obtained for each method. Performance was compared with manufacturer claims. Four of the tested platforms demonstrated a clear high-dose hook effect, while the other methods showed no hook effect at the highest level tested. Our results indicate that the hook effect may occur in some hCG assays, although the risk of reporting falsely low results was in most cases at higher concentrations than those indicated in manufacturers' product information. Assay design plays a major role in its occurrence. Laboratories should be aware of the assay limitations in this regard.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.