Abstract
BackgroundThe possibility of clearing Cell-free Plasma Hemoglobin (CPH) from human plasma may appear attractive, especially when considering the noxious effects that CPH has on the immune function and the renal damage caused by its filtration. The existence of the so-called High Cut-Off (HCO) filters, possessing pores as big as 60 kDa, could potentially allow the clearance of the αβ dimers (31.3 kDa), the form in which the α2β2 hemoglobin tetramers (62.6 kDa) physiologically dissociate in plasma. We present herein the first reported case in which such an attempt was made.Case presentationThe patient was a 51-year-old man with hemolytic crisis due to glucose-6-phosphate dehydrogenase deficiency, further complicated by pigment-induced nephropathy. He underwent a 48-h CVVHD session, in which a HCO filter was used. The Sieving Coefficient (SC) for CPH was initially 0.08 and decreased to 0.02 after 24 h. This unexpected low SC was due to the initial high concentration of CPH (4.24 g/L). At such concentrations, the α2β2 tetramer poorly dissociates into the αβ dimer; but increases exponentially at concentrations lower than 1 g/L.ConclusionsClearance of CPH through a HCO filter is technically feasible but its performance markedly relies on the initial concentration of CPH. Critically ill patients with smoldering hemolysis, as it happens during septic shock or ECMO treatment, may benefit the most from the use of this membrane in order to clear CPH.
Highlights
The possibility of clearing Cell-free Plasma Hemoglobin (CPH) from human plasma may appear attractive, especially when considering the noxious effects that CPH has on the immune function and the renal damage caused by its filtration
Clearance of CPH through a High Cut-Off (HCO) filter is technically feasible but its performance markedly relies on the initial concentration of CPH
Ill patients with smoldering hemolysis, as it happens during septic shock or Extracorporeal Membrane Oxygenation (ECMO) treatment, may benefit the most from the use of this membrane in order to clear CPH
Summary
Clearance of CPH through a HCO filter is technically feasible but its performance markedly relies on the initial concentration of CPH. Critically ill patients with smoldering hemolysis, as it happens during septic shock or ECMO treatment, may benefit the most from the use of this membrane in order to clear CPH. Keywords: Free hemoglobin, High-cut off filter, Acute kidney injury, Sepsis, Continuous renal replacement therapy
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.