Fast and automated image quality assessment (IQA) for diffusion MR images is a crucial step for swiftly making a rescan decision during or after the scanning session. However, learning a model for this task is challenging as the number of annotated data is limited and the annotation labels might not always be correct. As a remedy, we will introduce in this paper an automatic image quality assessment (IQA) method based on hierarchical non-local residual networks for pediatric diffusion MR images. Our IQA is performed in three sequential stages, i.e., 1) slice-wise IQA, where a nonlocal residual network is first pre-trained to annotate each slice with an initial quality rating (i.e., pass/questionable/fail), which is subsequently refined via iterative semi-supervised learning and slice self-training; 2) volume-wise IQA, which agglomerates the features extracted from slices of a volume, and uses a nonlocal network to annotate the quality rating for each volume via iterative volume self-training; and 3) subject-wise IQA, which ensembles the volumetric IQA results to determine the overall image quality pertaining to a subject. Experimental results demonstrate that our method, trained using only samples of modest size, exhibits great generalizability, and is capable of conducting rapid hierarchical IQA with near-perfect accuracy.