Abstract

The authors have reviewed the pattern of drainage of Hemovac devices in 107 arthroplasties of the hip. Drainage ceased in 55 hips (51%) within 18 hours and 93 hips (89%) within 24 hours. The mean drainage volume was 303 cc in femoral head replacement, 498 cc in simple total hip replacement, and 663 cc in extensive total hip replacement. Ninety nine per cent of the total drainage had been achieved within 24 hours, indicating that drainage devices can be removed then rather than between 24 and 72 hours as had been previously recommended. This early removal will help reduce the possibility of retrograde infection, and promote more comfortable rehabilitation after operation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.