Abstract
The hemostatic alterations in two adult patients who were supported by left ventricular assist devices (LVADs) because of postcardiotomy heart failure were evaluated. In both patients, fibrinopeptide A and thrombin-antithrombin III complex increased markedly during the first several days, and thereafter decreased moderately but remained above normal over the entire procedure. Furthermore, fibrinopeptide B beta 15-42 and alpha 2 plasmin inhibitor-plasmin complex were also markedly increased over the entire course of LVAD treatment. These data show that the LVAD system strongly activates both the coagulation and the fibrinolytic system, even when thromboembolic or bleeding complications are not clinically evident. Furthermore, the decrease in physiological coagulation inhibitors, especially protein C, indicates that these factors are activated and consumed during LVAD treatment. Because protein C is important in regulating the coagulation cascade during LVAD treatment, exhaustion of this system might result in thromboembolic complications during LVAD treatment.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.