Abstract
ObjectivesThis study aimed to compare hemodynamic characteristics of different self-expanding (SE) and balloon-expandable (BE) transcatheter heart valves (THV) in relation to native aortic annulus anatomy.BackgroundA patient centered THV selection becomes increasingly important as indications for transcatheter aortic valve replacement (TAVR) are extended towards lower risk populations.MethodsHemodynamic parameters including mean gradient (MG), effective orifice area (EOA), Doppler velocity index (DVI), degree of paravalvular regurgitation (PVR) and patient-prosthesis mismatch (PPM) were compared by valve type, label size and in relation to quintiles of native aortic annulus area.Results2609 patients were treated at 3 centers in Germany with SAPIEN 3 (n = 1146), ACURATE Neo (n = 649), Evolut R (n = 546) or Evolut Pro (n = 268) THV. SE THVs provided superior hemodynamics in terms of larger EOA, higher DVI and lower MG compared to BE THV, especially in patients with small aortic annuli. Severe PPM was less frequent in SE treated patients. The rate of PVR ≥ moderate was comparable for SE and BE devices in smaller annular dimensions, but remarkably lower for BE TAVR in large aortic annular dimensions (> 547.64 mm2) (2% BE THV vs. > 10% for SE THV; p < 0.001).ConclusionsPatients with small aortic annular dimensions may benefit hemodynamically from SE THV. With increasing annulus size, BE THV may have advantages since PVR ≥ moderate occurs less frequently.Graphical abstract