Abstract
IntroductionControlling temperature generation during implant site preparation is important to prevent implant early failure.AimThe present in vitro study aimed at measuring temperature variation generated during the initial osteotomy using both rotatory and piezo-surgical inserts.MethodsNine groups were defined according to drill and insert type, cooling volume (mL/min) and cooling temperature. A total of 315 implant site preparations were performed in an artificial bone sample and the temperature was measured using an infrared camera. Drills’ wear was assessed using scanning electron microscopy at baseline and after 10 and 35 utilizations.ResultsPiezo-surgical insert groups determined a temperature increase that was significantly higher than the one generated by rotatory drills groups (p < 0.001). When considering rotatory drills groups a temperature ≥ 40 °C was never recorded.ConclusionLower saline temperature implied a significant temperature decrease (p < 0.001), while the increase in cooling volume did not imply a temperature decrease. The scanning electron microscopy analysis of the drills demonstrated that little drill wear occurred up to 35 utilization times.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.