Abstract

In this work a 21 MHz radiofrequency (RF) exposure system was designed, constructed and evaluated as a testing platform for MRI device safety. The exposure system consisted of a birdcage coil powered in quadrature by a power system consisting of a hybrid coupler, a power splitter, a phase shifter and two baluns. The birdcage coil was designed as a low-pass 16 rung coil. Both the coil and the power system were fabricated as printed circuit boards with a copper thickness of 205.7 μm and a G10 substrate thickness of 581.7 μm. The birdcage was assembled by wrapping the copper clad G10 inside of an acrylic tube. A shield for the coil was created as a cylindrical ground plane with a diameter of 39.7 cm. The birdcage and the power system were simulated in ADS and EMPro, Keysight Technologies Inc. both as lumped elements and using full wave 3D simulations. Simulation results were compared to measurements of the produced components to validate each component and entire system. In addition, a 3D magnetic field probe was developed to measure and compare the fields to analytical and numerical calculations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.