Abstract

Implantable cardiac pacemakers are medical devices that can monitor and correct abnormal heart rhythms. To provide the necessary safety assurance for pacemaker software, both testing and verification of the code, as well as testing the entire pacemaker hardware in the loop, is necessary. In this paper, we present a hardware testbed that enables detailed hardware-in-the-loop simulation and energy optimisation of pacemaker algorithms with respect to a heart model. Both the heart and the pacemaker models are encoded in Simulink/Stateflow™ and translated into executable code, with the pacemaker executed directly on the microcontroller. We evaluate the usefulness of the testbed by developing a parameter synthesis algorithm which optimises the timing parameters based on power measurements acquired in real-time. The experiments performed on real measurements successfully demonstrate that the testbed is capable of energy minimisation in real-time and obtains safe pacemaker timing parameters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.