Abstract

Low-cost, rapid, and accurate acquisition of minimum inhibitory concentrations (MICs) is key to limiting the development of antimicrobial resistance (AMR). Until now, conventional antibiotic susceptibility testing (AST) methods are typically time-consuming, high-cost, and labor-intensive, making them difficult to accomplish this task. Herein, an electricity-free, portable, and robust handyfuge microfluidic chip was developed for on-site AST, termed handyfuge-AST. With simply handheld centrifugation, the bacterial-antibiotic mixtures with accurate antibiotic concentration gradients could be generated in less than 5 min. The accurate MIC values of single antibiotics (including ampicillin, kanamycin, and chloramphenicol) or their combinations against Escherichia coli could be obtained within 5 h. To further meet the growing demands of point-of-care testing, we upgraded our handyfuge-AST with a pH-based colorimetric strategy, enabling naked eye recognition or intelligent recognition with a homemade mobile app. Through a comparative study of 60 clinical data (10 clinical samples corresponding to six commonly used antibiotics), the accurate MICs by handyfuge-AST with 100% categorical agreements were achieved compared to clinical standard methods (area under curves, AUCs = 1.00). The handyfuge-AST could be used as a low-cost, portable, and robust point-of-care device to rapidly obtain accurate MIC values, which significantly limit the progress of AMR.