Abstract

There have been some reports about possible effect of the hand presence on the head SAR if hand phantom is included in the measurements of the head SAR compliance assessment procedure. The objective of this computational study was to examine the reported effect by using realistic head models and realistic CAD based phone models. A commercially available FDTD based EM solver was used to carry out the computational work. Based on the results of this study considering the SAR values without hand phantom as reference, following conclusions can be made: 1. In general presence of the hand lead to significantly less conservative SAR values in the head for large majority of cases 2. For lower band GSM frequencies the presence of the hand decreases the head SAR up to ~70%. 3. For the upper band GSM frequencies the presence of the hand decreases the head SAR up to ~55%. Based on the results of this study the present SAR compliance protocol where hand phantom is not included leads to more conservative head SAR results compared to the cases where hand is included.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.