Abstract

A 92-year-old woman underwent an implantation of a leadless pacemaker (Micra; Medtronic, Inc, Minneapolis, MN) for complete atrioventricular block after a transvenous lead extraction due to a pocket infection of a dual chamber pacemaker. Marked scoliosis and a humpback due to an advanced age made it impossible to direct the tip of the pacemaker delivery catheter towards the right ventricular septum or apex and shape the catheter into a gooseneck-shape. Thus, by attaining a halo-catheter shape of the delivery catheter, the catheter tip could be directed toward the infero-basal portion of the right ventricular septum. The pacemaker was successfully deployed at that site without any complications, and good device parameters were achieved. The halo-shape technique may be also an alternative method for delivering a leadless pacemaker in patients with an unsuccessful delivery of a leadless pacemaker to the right ventricular septum using the conventional gooseneck-shape technique.