Abstract

We evaluated five commercially available glare testers to determine how accurately each device predicted outdoor Snellen acuity in cataract patients. Predicted outdoor Snellen acuity was compared to actual Snellen outdoor acuity for the Miller-Nadler glare tester, the brightness acuity tester (BAT), the InnoMed true vision analyzer (TVA), the VisTech VCT 8000, and the EyeCon 5. All five devices were evaluated in a single series of eyes. We ranked the five glare testers as follows based upon their ability to predict outdoor Snellen acuity within one Snellen line: (1) BAT (73%), (2) TVA (69%), (3) VisTech VCT 8000 (56%), (4) Miller-Nadler (47%), and (5) EyeCon 5 (15%). All five devices were also evaluated for their capacity to provide false negative or false positive results and for testing time per eye, test difficulty level for the patient, versatility, upgrading capacity, examining room adaptability, and price. We concluded that glare testing of cataract patients is necessary to measure visual disability and that glare testing devices vary in their ability to predict outdoor vision.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.