Abstract

The disparity between functional outdoor vision and the acuity measured in the standard refracting lane is well known among clinicians. A simple device, the brightness acuity tester (B.A.T.), was developed to predict a patient's functional outdoor acuity. The B.A.T. has an illuminated hemispheric bowl, 60 min in diameter, with a 12-mm aperture. Fourteen normas and 50 patients with cataracts were tested using the B.A.T. and then tested outside in bright sunlight. The B.A.T. correlated extremely well (r = +0.84, P <.0001) with the acuities measured outside. There was no decrease in visual acuity in the 14 normal patients, but there was a one to ten line decrease in vision among the cataract patients. Upon retesting, the B.A.T. scores did not vary, while outside testing scores did change due to variable outdoor ambient, light levels on sunny days. The B.A.T. is a simple, repeatable, useful test for predicting functional outdoor acuity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.