Abstract

To determine whether dimensional scaling (relative to the human) is necessary for screwed pins used in externally applied fracture fixation studies on sheep, geometrical data were determined for six ovine tibiae. Each tibia was potted relative to a lengthwise reference axis and sectioned at 5 per cent length intervals over its central 80 per cent. Enlarged (280 per cent) images of each cross-section were digitized at 1 mm increments around the periphery of the periosteal and endosteal surfaces, the data were digitally filtered, and geometrical properties were computed to include cross-sectional area A, maximum and minimum second moments of area (Imax and Imin), polar second moment of area J, and effective polar second moment of area J(eff). Proportional scaling of geometrical properties with respect to bone length (L2 for A, and L4 for second moments of area) significantly (p < 0.000001) decreased the coefficient of variation in data by an average 36 per cent. From 30-90 per cent distal, J(eff) for the ovine tibia is smaller but within 7 per cent of J--in stark contrast with the human tibia, where J(eff) has been reported as 70-80 per cent of J over the same tibial length. While previous ovine studies involving external fixator pins have employed the same diameter of pin as has been used in humans (that is 5 or 6 mm), a 'first-order' approximation of the data for A, Imax, Imin and J(eff) suggests these pins should be scaled down to 4 mm and 4.75 mm respectively for use on the ovine tibia over the range 25-80 per cent distal along its length.