Abstract

AbstractCapillary columns of 0.3–0.35 mm internal diameter and 0.3–7.7 m length, packed with 3 to 30 μm octadecylsilica stationary phases as used for liquid chromatography, were applied to gas chromatographic separation of low boiling hydrocarbons. Van Deemter plots for these columns showed the optimum column efficiency to occur at linear velocities of 4–5 cm/s. A short column was applied to the rapid separation of components of a natural gas and impurities in standard gases, while a long column was applied to the separation of complex mixtures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.