Abstract
Measurement of arterial wall thickness is an integral component of vascular properties and health assessment. State-of-the-art automated or semi-automated techniques are majorly applicable to B-mode images and are not available for entry-level in-expensive devices. Considering this, we have earlier developed and validated an image-free (A-mode) ultrasound device, ARTSENS® for the evaluation of vascular properties. In this work, we present a novel gaussian-mixture modeling-based method to measure arterial wall thickness from A-mode frames, which is readily deployable to the existing technology. The method's performance was assessed based on systematic simulations and controlled phantom experiments. Simulations revealed that the method could be confidently applied to A-mode frames with above-moderate SNR (>15 dB). When applied to A-mode frames acquired from the flow-phantom setup (SNR > 25 dB), the mean error was limited to (2 ± 1%), and RMSE was 19 μm, on comparison with B-mode measurements. The measured and reference wall thickness strongly agreed with each other (r = 0.88, insignificant mean bias = 7 μm, p = 0.16). The proposed method was capable of performing real-time measurements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.