Abstract

BackgroundECG-gated cardiac CT is now widely used in infants with congenital heart disease (CHD). Deep Learning Image Reconstruction (DLIR) could improve image quality while minimizing the radiation dose. ObjectivesTo define the potential dose reduction using DLIR with an anthropomorphic phantom. MethodAn anthropomorphic pediatric phantom was scanned with an ECG-gated cardiac CT at four dose levels. Images were reconstructed with an iterative and a deep-learning reconstruction algorithm (ASIR-V and DLIR). Detectability of high-contrast vessels were computed using a mathematical observer. Discrimination between two vessels was assessed by measuring the CT spatial resolution. The potential dose reduction while keeping a similar level of image quality was assessed. ResultsDLIR-H enhances detectability by 2.4% and discrimination performances by 20.9% in comparison with ASIR-V 50. To maintain a similar level of detection, the dose could be reduced by 64% using high-strength DLIR in comparison with ASIR-V50. ConclusionDLIR offers the potential for a substantial dose reduction while preserving image quality compared to ASIR-V.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.