Abstract

We describe a vacuum apparatus for determining the outgassing rate into vacuum, the diffusion coefficient, and the amount of gas absorbed for various materials. The diffusion coefficient is determined from a model applied to time-dependent desorption data taken using a throughput method. We used this method to determine the diffusion coefficient, D, for H2O in 3-D printed acrylonitrile butadiene styrene (ABS). We found DH2O = 8.3 × 10-8 cm2/s ± 1.3 × 10-8 cm2/s (k = 1; 67% confidence interval) at 23.2 °C. This result was compared to the diffusion coefficient determined another by a gravimetric method, in which the sample weight was monitored as it absorbed gas from the atmosphere. The two methods agreed to within 3%, which is well within the uncertainty of the measurement. We also found that at least 80% of the atmospheric gas (air) absorbed by the ABS is water. The total amount of all atmospheric gas absorbed by ABS was about 0.35% by weight when exposed to ambient air in the laboratory, which was at a pressure of 101 kPa with a relative humidity of 57% at 22.2 °C.