Abstract
Proportional counters filled with tissue equivalent gas mixtures (TEPC) can be used to simulate interactions and energy transferred to small tissue volumes. One criteria which allows to use TEPC as the dose meter is that the particle ranges are larger compared to the gas volume. TEPC achieve this by operating at low gas pressures. Single ionization events dominate the distribution of low-LET radiation at low gas pressure and therefore their detection is of primary importance, a high gas gain is necessary. Therefore gas gain factor has been measured for Methane- and Propane-based tissue equivalent gas mixtures. The highest stable gas gains, second ionization Townsend coefficient and electron avalanche dimensions have been determined.