Abstract
An FIA conductometric method for monitoring of ammonium ions in river and waste waters is proposed. A very good detection limit (xD 5.6 μmol l-1) was obtained at 100/20 s preconcentration/rinsing times and 2 ml sample consumption per analysis at the flow rate 0.8 ml min-1. The detection limit xD can be further improved at longer preconcentration/ rinsing times. Precipitation of polyvalent metal ions (mainly Fe(III)) was eliminated by addition of 0.5 mmol l-1 EDTA prior to injection. The results are comparable with those obtained by spectrophotometry (Nessler's reagent).