Abstract

AbstractIt is demonstrated that linear injection characteristics are obtained for a wide boiling point range sample using a temperature‐programmed injector in combination with wide‐bore fused silica columns. The applicability of the described combination for high temperature simulated distillation is described. The method, using external standardization, gives accurate and repeatable results for different types of samples in the boiling range between 50 and 750°C. The lifetime of the fused silica wide‐bore columns was found to be acceptable, viz. over 80 temperature‐programmed cycles between ambient and 430°C. Some comments are made on the accuracy of boiling points for normal alkanes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.