Abstract

The detection and quantitation of slight increases of plasma homocysteine levels is of growing interest. This has prompted us to develop a highly sensitive and accurate capillary gas chromatography–mass spectrometry (GC–MS) method. The method proved to be highly sensitive (DL=0.17 μmol/l) with between- and within-run precision less than 6% and 7%, respectively. Reference values of plasma total homocysteine have been determined for men ( n=39) and women ( n=36), showing a significant difference ( P=0.003) between gender. Preliminary results in cerebrovascular accidents and in venous thrombosis are presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.