Purpose: This study provides a simple protocol for validation of the gamma passing rates and to identify the optimum values of % dose and mm criteria for dose distributions measured with a detector array. Methods: We chose ArcCHECK detector array to illustrate the concepts. We used plans with uniform or quasi-uniform dose distributions along the detector array for testing in the presence of dose errors. For testing sensitivity to spatial shift we employed a plan with approximately constant dose gradient along the axis of the instrument. Results: We identified a representative set of parameters which describe performance of a detector array. We determined the minimum gamma-index acceptance criteria allowing the passing rates to reach 100% in the absence of errors, and identified the minimum fully detectable errors for such criteria. For our baseline plans delivered to ArcCHECK, 100% passing rates were obtained for 1.5% dose criterion together with ±3% minimum error detectable at 100% rate, and for 1.5 mm criterion together with the minimum fully detectable error of ±3 mm. We inspected the impact of selected program options on the passing rates. Conclusions: The protocol we developed provides a simple method of commissioning-style analysis of a detector array without a need for analysis of a large number of clinical plans.

Highlights

  • A number of detector arrays have been developed for quality assurance (QA) of complex radiation treatment plans

  • The passing rates as a function of the % dose criterion in the gamma index for ArcCHECK irradiated with the uniform-dose plan are shown in Figure 3 as black circles

  • 100% of the detectors should pass at any non-zero % dose criterion, and smaller passing rates are a reflection of uncertainties in the system

Read more

Summary

Introduction

A number of detector arrays have been developed for quality assurance (QA) of complex radiation treatment plans. Early-generation systems contained a set of detectors arranged in a line, e.g. Profiler 2, (Sun Nuclear Inc., Melbourne, FL) or in a plane, e.g. MatriXX (IBA Dosimetry GmbH, Schwarzenbruck, Germany). How to cite this paper: Szpala, S. and Kohli, K. (2015) Gamma-Index Passing Rates in Baseline Plans Measured with a Detector Array. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 4, 326-337. Kohli struments evolved to allow measuring the dose distribution in 3D space. Delta 4 (ScandiDos Inc., Uppsala, Sweden) consists of two planes of detectors arranged in “X” configuration and ArcCHECK (Sun Nuclear Inc., Melbourne, FL) consists of an array of detectors located on a cylindrical surface

Objectives
Methods
Results
Discussion
Conclusion