Abstract

The healing process of bone fractures can be monitored by a measurement of the osteosynthesis plate bending. An electrical resonant circuit consisting of a coil with a magnetostrictive Galfenol core and a capacitance enables an indirect wireless measurement of the plate bending. The Galfenol core is manufactured with a thin film technology using a Galfenol Fe <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">83</sub> Ga <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">17</sub> alloy. The sensor has a quit good linear resonant frequency-force characteristic. Both, time and frequency domain measurement techniques for the sensor, use an external measurement coil. In the frequency domain the transformed sensor impedance is evaluated and the resonance frequency is determined by a local extremum. In the time domain a short energizing sinus pulse is transmitted, the sensor response is received and a frequency counting, Fourier or Wavelet technique is used for resonance frequency detection.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.