Abstract

Abstract PADC detectors are commonly used both as radon detectors and as personal dosemeters for neutron radiation. In both cases the measurement in workplace fields are characterized by long term irradiations, lasting up to several months. During this period the detectors undergo to uncontrolled and unmonitored environmental conditions. In a recent work the same authors demonstrated that the environmental conditions, mainly temperature, can seriously affect the detector sensitivity. The temperature effect is to reduce the V ratio by decreasing the track etching velocity V t . This causes a decrease in limit angle and efficiency. This paper describes the same ageing/fading test made with CR39 from a different supplier. In this case it seems that the temperature has a negligible effect, so no compensation algorithm is needed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.