Abstract

The higher value of uncertainty in the flow measurement signal from transit-time clamp-on ultrasonic flowmeters (USFMs) compared with transit-time spool-piece USFMs is due to several factors, some of which have been considered in this paper. However, the main focus of this paper has been on the effect of separation distance between the transducers on the flowmeter output signal. The sources of experimental error that might affect the validity of the data obtained for the transducer separation are examined, and then, use of ray-tracing and finite-difference models is made to examine this relationship theoretically.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.