Abstract
A fully automated technique for high-performance liquid chromatographic analysis of whole blood and plasma is described. Samples are automatically injected into a dialyser where proteins and blood cells are removed. The dialysates are concentrated on a small column prior to analysis. This technique is used for the determination of oxytetracycline in whole blood and plasma. After dialysis oxytetracycline and the internal standard, tetracycline, are retained on a polystyrene enrichment column and subsequently separated on a polystyrene analytical column by ion-pair chromatography. Using ultraviolet detection 50 ng/ml can be detected. Validation showed good within-day and between-day accuracy and precision. Different oxytetracycline concentrations were found in plasma and whole blood. This difference varied between the species.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Chromatography B: Biomedical Sciences and Applications
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.